ASTM F1160 Shear and Bending Fatigue Testing of Calcium Phosphate and Composite Calcium Phosphate/Metallic Coatings
The procedure for determining the shear and bending fatigue performance of calcium phosphate coatings and porous and nonporous metallic coatings, as well as the bending fatigue performance of metallic coatings over sprayed with calcium phosphate, is covered by ASTM F1160. The SI units are to be taken as standard.

TRUSTED BY




Precision-driven testing for dimensional accuracy and compliance
- Overview
- Scope, Applications, and Benefits
- Test Process
- Specifications
- Instrumentation
- Results and Deliverables
Overview
ASTM F 1160 shear and bending fatigue testing is the standardized test method for evaluating the fatigue performance of calcium phosphate and composite calcium phosphate/metallic coatings applied to metallic substrates — measuring the resistance of biomedical coating systems to shear and bending fatigue failure under defined cyclic loading conditions. ASTM F 1160 defines the specimen preparation, fatigue loading configurations, and failure assessment procedures required to characterize coating fatigue performance — providing biomedical engineers, implant manufacturers, and coating developers with reliable fatigue data for coating selection, implant design validation, and supplier assessment programs.

Scope, Applications, and Benefits
Scope
ASTM F 1160 evaluates the shear and bending fatigue performance of calcium phosphate and composite coatings by subjecting coated metallic specimens to defined cyclic shear and bending load conditions, assessing coating adhesion retention, crack initiation, delamination, and fatigue failure across specified load cycles and stress amplitudes for biomedical implant coating systems.
ASTM F 1160 shear and bending fatigue testing evaluates:
- Shear fatigue resistance of calcium phosphate coatings on metallic implant substrates
- Bending fatigue performance of composite calcium phosphate/metallic coating systems
- Coating adhesion retention under cyclic shear and bending loading conditions
- Crack initiation, propagation, and delamination behavior of biomedical coatings under fatigue
- Compliance with ASTM F 1160 fatigue performance requirements for implant coating systems
Applications
- Hydroxyapatite and calcium phosphate-coated orthopedic implant components
- Composite calcium phosphate/titanium and calcium phosphate/metallic coating systems
- Dental implant and prosthetic component coating fatigue performance assessment
- Spinal fusion device and joint replacement implant coating fatigue verification programs
- Biomedical coating developers requiring ASTM F 1160 shear and bending fatigue test data
Benefits
- Provides reliable shear and bending fatigue data for biomedical coating selection and design
- Supports implant coating specification compliance and supplier assessment programs
- Identifies coating fatigue deficiencies before implant production and clinical deployment
- Delivers traceable fatigue test records for engineering and regulatory submissions
- Reduces implant coating failure risk by verifying fatigue performance early in the approval cycle
Test Process
Sample Preparation
Calcium phosphate coated metallic specimens prepared per ASTM F 1160 geometry requirements.
1Fatigue Load Application
Cyclic shear or bending loads applied at defined stress amplitude and frequency per ASTM F 1160.
2Failure Monitoring
Specimens monitored throughout cycling for coating delamination, cracking, and fatigue failure.
3Data Analysis & Reporting
Fatigue performance assessed against ASTM F 1160 acceptance criteria for test compliance result.
4Technical Specifications
| Parameter | Details |
|---|---|
| Applicable Coatings | Calcium phosphate, hydroxyapatite, and composite calcium phosphate/metallic coating systems |
| Specimen Geometry | Defined metallic substrate dimensions per ASTM F 1160 shear and bending configuration |
| Cyclic Loading Conditions | Defined stress amplitude and frequency at 1 Hz to 30 Hz per ASTM F 1160 fatigue requirements |
| Test Environment | Physiological saline solution at 37°C ± 1°C per ASTM F 1160 simulated implant conditions |
| Measured Outputs | Fatigue life cycles, delamination observations, coating integrity data, and test compliance result |
Instrumentation Used for Testing
- Servo-hydraulic or electromagnetic fatigue testing machine with defined load control
- Physiological saline solution bath at 37°C for simulated implant environment testing
- Precision load cell and displacement transducer for cyclic load measurement
- Optical and scanning electron microscopy for coating delamination and failure analysis
- Specimen preparation equipment for defined substrate geometry and coating application
- Data acquisition and fatigue cycle counting system
Results and Deliverables
- Cyclic load and displacement data throughout the shear and bending fatigue test periods
- Fatigue life cycle data and coating failure observations per ASTM F 1160 conditions
- Coating delamination, crack initiation, and fatigue failure mode observations
- Test compliance result assessed against ASTM F 1160 implant coating fatigue requirements
- ASTM F 1160 fatigue test report
Partnering with Infinita Lab for Optimal Results
Infinita Lab addresses the most frustrating pain points in the ASTM F1160 testing process: complexity, coordination, and confidentiality. Our platform is built for secure, simplified support, allowing engineering and R&D teams to focus on what matters most: innovation. From kickoff to final report, we orchestrate every detail—fast, seamlessly, and behind the scenes.
Looking for a trusted partner to achieve your research goals? Schedule a meeting with us, send us a request, or call us at (888) 878-3090 to learn more about our services and how we can support you. Request a Quote
Frequently Asked Questions
ASTM F1160 includes cyclic loading frequency, load amplitude, specimen geometry, coating thickness, and number of cycles, evaluating fatigue performance of calcium phosphate and composite coatings under shear and bending stresses.
ASTM F1160 applies repeated cyclic shear and bending loads at controlled frequencies, simulating physiological conditions to evaluate durability and adhesion of coatings on metallic substrates.
ASTM F1160 is commonly applied to calcium phosphate coatings, hydroxyapatite coatings, and composite coatings deposited on metallic implants used in biomedical and orthopedic applications.
ASTM F1160 evaluates coating fatigue life, crack initiation, delamination behavior, and failure mechanisms under cyclic loading to assess long-term performance of coated biomedical implants.
ASTM F1160 evaluates fatigue behavior under controlled laboratory conditions and may not fully represent in vivo environments where biological interactions, corrosion, and complex loading conditions affect coating performance.
Case Studies
In-depth examination of genuine material testing solutions
Dopant and Ultra-Low Concentration Elemental Analysis Using STEM…
Introduction to STEM-EELS for Elemental Analysis Scanning Transmission Electron Microscopy (STEM) combined with Electron Energy Loss...
Read Case StudyAnalysis of PVC Pipe Degradation Using FTIR Spectroscopy
PVC Pipe in Infrastructure — and Why Degradation Matters Polyvinyl chloride (PVC) pressure pipe is one...
Read Case StudyNano-scale roughness measurement of Si-wafers by Atomic Force…
Nano-scale surface roughness is a critical parameter in fabricated thin-films that are used in optics, solar...
Read Case Study
Request a Quote
Submit your material details and receive testing procedures, pricing, and turnaround time within 24 hours.