ASTM E967 Temperature Calibration Testing for DSC & DTA Instruments

ASTM E967 outlines the procedure for temperature calibration of differential thermal analyzers and differential scanning calorimeters across a temperature range from −40 to +2500°C. All values given in the standard are in SI units.

    Talk to an Expert
    ASTM E967 Temperature Calibration Testing for DSC & DTA Instruments

    TRUSTED BY

    Precision-driven testing for dimensional accuracy and compliance

    • Overview
    • Scope, Applications, and Benefits
    • Test Process
    • Specifications
    • Instrumentation
    • Results and Deliverables

    Overview

    ASTM E 967 Temperature Calibration Testing: The standardized testing procedure to calibrate the temperature scale of differential scanning calorimeters and differential thermal analyzers using certified reference materials to ensure traceable accuracy in the range of temperatures that can be achieved in a defined melting point calibration procedure. The reference materials to be used in the calibration process, as well as the procedure for correcting the accuracy of DSCs and DTAs, are defined in the ASTM E 967 testing procedure to ensure accurate data can be obtained in DSC/DTA testing.

    Scope, Applications, and Benefits

    Scope

    ASTM E 967 standards are used to calibrate the temperature scale of DSC and DTA instruments by determining the onset melting points of certified reference standards within the equipment’s operating range. The onset melting points are compared to the certified reference standards to establish the temperature correction factors and accuracy of the equipment temperature scale.

    ASTM E 967 temperature calibration testing evaluates:

    • Temperature scale accuracy of DSC and DTA instruments across the operating temperature range
    • Onset melting temperature measurement of certified reference materials per ASTM E 967
    • Temperature correction factor determination for DSC and DTA instrument calibration
    • Instrument temperature linearity and accuracy across defined calibration temperature points
    • Compliance with ASTM E 967 temperature accuracy acceptance criteria for thermal analysis instruments

    Applications

    • DSC and DTA instrument temperature calibration for polymer thermal analysis programs
    • Pharmaceutical DSC temperature calibration for melting point and purity assessment
    • Materials research DSC and DTA temperature accuracy verification programs
    • Production laboratory thermal analysis instrument temperature calibration programs
    • Calibration laboratories requiring ASTM E 967 DSC and DTA temperature calibration data

    Benefits

    • Provides reliable DSC and DTA temperature calibration data for thermal analysis accuracy assurance
    • Supports thermal analysis instrument compliance and measurement traceability requirements
    • Identifies temperature accuracy deficiencies in DSC and DTA instruments before analysis programs
    • Delivers traceable temperature calibration records for laboratory and regulatory submissions
    • Reduces thermal analysis measurement risk by verifying instrument temperature accuracy early in calibration

    Test Process

    Reference Material Preparation

    Certified reference material specimens encapsulated in DSC or DTA pans per ASTM E 967 requirements.

    1

    Thermal Scanning

    Reference specimens heated through melting transition at 10°C/min scan rate per ASTM E 967.

    2

    Onset Temperature Measurement

    Melting onset temperature determined from extrapolated leading edge of melting endotherm peak.

    3

    Data Analysis & Reporting

    Measured onset temperatures compared against certified values to determine temperature correction and test compliance result.

    4

    Technical Specifications

    ParameterDetails
    Applicable InstrumentsDifferential scanning calorimeters and differential thermal analyzers across defined operating ranges
    Reference MaterialsCertified NIST-traceable reference materials including indium 156.6°C, tin 231.9°C, and zinc 419.5°C
    Heating Rate10°C/min standard heating rate per ASTM E 967 calibration scan conditions
    Temperature AccuracyOnset temperature accuracy within ±0.5°C of certified reference material melting temperature
    Measured OutputsOnset temperatures, temperature correction factors, calibration curves, and test compliance result

    Instrumentation Used for Testing

    • A differential scanning calorimeter or differential thermal analyzer requires temperature calibration
    • Certified NIST-traceable reference materials ,including indium, tin, and zinc standards
    • Precision analytical balance for reference material specimen mass measurement
    • Aluminum and hermetic DSC pan encapsulation equipment
    • Specialist DSC and DTA thermal data acquisition and onset temperature analysis software
    • Data reporting and calibration curve visualization system

    Results and Deliverables

    • Onset melting temperature measurements for all certified reference materials per ASTM E 967
    • Temperature correction factor data across the instrument calibration temperature range
    • Calibration curve showing measured versus certified reference material temperatures
    • Test compliance result assessed against ASTM E 967 ±0.5°C temperature accuracy requirements
    • ASTM E 967 DSC and DTA temperature calibration report

    Why Choose Infinita Lab for ASTM E 967?

    Infinita Lab offers comprehensive ASTM E 967 testing services, a Comprehensive lab network, project management, confidentiality, and rapid turnaround. Trust Infinita Lab for your material testing needs, Faster test results, cost savings, and reduced administrative workload.

    Looking for a trusted partner to achieve your research goals? Schedule a meeting with us, send us a request, or call us at (888) 878-3090  to learn more about our services and how we can support you. Request a Quote

    Frequently Asked Questions

    ASTM E967 uses reference materials with known transition temperatures, controlled heating rates typically 1 to 20°C per minute, baseline stability, and peak temperature accuracy to calibrate DSC and DTA instruments.

    ASTM E967 uses standard materials such as indium, tin, zinc, and aluminum with well-defined melting points to verify and adjust temperature accuracy of thermal analysis instruments.

    ASTM E967 calibration covers a broad temperature range depending on selected reference materials, typically from near ambient conditions up to several hundred degrees Celsius for accurate instrument performance verification.

    ASTM E967 is used to calibrate differential scanning calorimeters and differential thermal analyzers, ensuring accurate temperature measurement during thermal analysis of materials.

    ASTM E967 calibration accuracy depends on proper sample preparation, heating rates, and instrument condition, and may not account for variations during real testing conditions or different sample types.

     Request a Quote

    Request a Quote

    Submit your material details and receive testing procedures, pricing, and turnaround time within 24 hours.



    • ddd
      Quick Turnaround and Hasslefree process
    • ddd
      Confidentiality Guarantee
    • ddd
      Free, No-obligation Consultation
    • ddd
      100% Customer Satisfaction
    Home / services / ASTM E967 Temperature Calibration Testing for DSC & DTA Instruments

    Discover more from Infinita Lab

    Subscribe now to keep reading and get access to the full archive.

    Continue reading

    ×

    Talk to an Expert

      Connect Instantly

      (888) 878-3090
      Ensure Quality with the Widest Network of Accredited Labs
      • ddd
        Quick Turnaround and Hasslefree process
      • ddd
        Confidentiality Guarantee
      • ddd
        Free, No-obligation Consultation
      • ddd
        100% Customer Satisfaction

        ddd

        Start Material Testing